HIT Box: Quick Hit Clinical Demo (sponsored by Sig ...
HIT Box
HIT Box
Back to course
[Please upgrade your browser to play this video content]
Video Transcription
Video Summary
To verify a hearing aid is performing to standards, first remove it from the client and save any wearer settings if needed. Reconfigure the device to test settings in the programming software, then place it correctly in the test box and coupler. Use the reference microphone positioned near the hearing aid microphone. Run ANSI measurements at full-on gain and reference test gain to check outputs like max OSPL90, average gain, frequency range, and distortion. Compare results with the manufacturer’s data sheet, then restore the original settings afterward.
Keywords
hearing aid verification
ANSI measurements
test box and coupler
OSPL90
manufacturer data sheet
×